Sunpump offers the CPA current preamplifier for THz power measurement systems. Combined with THZ10/THZ20/THZ30 THz detectors it forms a THz power meter for power measurement of cw and chopped radiation.
The CPA current preamplifier is required for power measurement of incoming radiation. It consists of a transimpedance amplifier IC at the input and two further voltage amplifier stages, plus components for noise reduction and offset regulation. Conversion factors between 10⁷ V/W and 10¹⁰ V/W can be realized.
• Connectors: BNC
• Amplification: 10⁷, 10⁸, 10⁹, 10¹⁰ V/A (4-step switch)
• Bandwidth: 50 Hz – 250 Hz, switchable
• Power supply: 5 V, Micro-USB
• Sensitivity: detector current sensitivity × amplifier gain
THz laser power measurement; cw laser power measurement with chopper; THz power meter with THZ10/20/30 detectors; lock-in pre-amplification.
| Connectors | BNC |
|---|---|
| Amplification | 10⁷ / 10⁸ / 10⁹ / 10¹⁰ V/A |
| Bandwidth | 50 Hz – 250 Hz, switchable |
| Power supply | 5 V, Micro-USB |
| Conversion factor | 10⁷ – 10¹⁰ V/W |
| Input type | Transimpedance amplifier (current input) |
Function: Power measurement of incoming radiation. Input stage is a transimpedance amplifier IC with two further voltage amplifier stages, plus noise reduction and offset regulation.
Sensitivity: Detector sensitivity × amplifier gain (e.g. detector 10⁻⁶ A/W and CPA 10⁹ V/A gives 1000 V/W).
Amplification: 4-step switch 10⁷...10¹⁰ V/A; bandwidth fixed 50 Hz or 250 Hz; custom adaptation on request.
Detection limit: Depends on amplification, bandwidth and detector diameter. With THZ10: 8 µW (17 Hz), 20 µW (70 Hz), 25 µW (200 Hz), 100 µW (4 kHz).